 |
|
  |
|
LITTON LITPAC II
The LITPAC II (figures
12-13) is a positive-pressure,compressed-gas (39% O2), closed-circuit
apparatus with a refillable, CO2-absorbent canister. It was certified by
NIOSH in 1994 as a 2-hr, positive-pressure apparatus. Its targeted market is
mainly the fire service and other users who requirepositive pressure even at
high work rates. These users desire durations longer than open-circuit
apparatus can provide, but do not want high O2 levels in the breathing
circuit. This apparatus was designed to pass the anticipated National Fire
Protection Association 1984 standard for closed-circuit apparatus. The two
gas cylinders contain 1,036 L ATPD of gas when filled to 4,500 psi. The
LITPAC II is a demand-only apparatus with approximately 20% of each breath
coming from the cylinders through a positive-pressure-biased demand valve
regulator. The same portion of exhaled air is vented out of the circuit on
every exhalation through a pressure-biased exhalation/relief valve. This
strategy spares the CO2-absorbent to some extent and provides cooler
breathing air, but requires a large quantity of supply gas. If one considers
that open-circuit apparatus draw 100% of each breath from the gas supply
versus approximately 4% for typical closed-circuit apparatus, this apparatus
can be seen as a cross between open- and closedcircuit. The user exhales
into the face mask, through the exhalation breathing hose, into an
exhalation/relief valve assembly containing a check valve and a relief
valve, both of which are spring-loaded. The exhaled air then either exits
the circuit via the relief valve or enters the scrubber inlet bag. The air
that enters the scrubber inlet bag then passes through the CO2- absorbent
canister into a scrubber outlet bag, then into the pressure vessel bag,
which encases the two compressed-gas cylinders. A small fourth bag made of
thicker material, functioning more as a duct, connects the pressure vessel
bag with the regulator. Inhalation draws air from the breathing bags and the
gas cylinders. The gas from the cylinders enters the regulator through an
injector nozzle, creating negative pressure in the regulator, which draws
air in from the breathing bags, functioning as a sort of power-assist. This
combination gascylinder/ breathing-bag air is then drawn through the
inhalation breathing hose and back into the face mask.
|
|