Orifices used in semi-closed
rebreathers
Author: Jan Willem
Bech
Publication:
http://www.therebreathersite.nl
February 2005
|
I wrote this article
because many questions were being asked in newsgroups about flow through an
orifice. A lot of people are developing manually controlled rebreathers
these days. I thought it would be interesting to understand why Dräger used
two orifices instead of one for the Dolphin rebreather. In the earlier
published article written by Jan Jahns, he explained the physical
backgrounds of orifices. This article tries to explain the practical use of
the theory. I would also like to shine some light on the results of a wrong
IP adjustment of the CMF first stage and what happens if one of the two
robins gets clogged in the dosage unit.
|
|
Many semi-closed
rebreathers are designed to work on the constant mass flow principle (CMF).
This kind of rebreather is also called an active semi-closed rebreather. The
functioning of the gas supply system in short means the following:
Gas from a cylinder is
reduced by a first stage regulator. The intermediate pressure is brought to
a very small part in which a restriction hole has been drilled. This
component is called an orifice. Gas laws stipulate that in certain
conditions the quantity of gas (mass) remains constant. We can therefore
inject a fixed mass of gas to the breathing loop of a rebreather.
Since at certain
exertion the human body uses the same amount
of oxygen at every depth, the quantity of oxygen we have to inject is not
depth-related.
The gas mixture
composition therefore is not determined by the amount of oxygen we need, but
by its partial pressure and by the narcotic effect of inert gasses such as
nitrogen. To diminish the narcotic effect of nitrogen and therefore the
decompression obligation, we may enlarge the oxygen dosage to a maximum
partial pressure of 1.6 bar (military limits 2.0 bar).
If we intend to exceed
the depth that is determined by nitrogen sensitivity, helium is to be added
(END).
How to choose the right
orifice for a certain gas mixture? How to design a gas dosage system of a
semi-closed rebreather?
Let’s start off with a
number of basic principles:
-
When dealing with a SC rebreather, a constant dose of Nitrox or Trimix
mixture is added to the loop. Only a part of oxygen in the mixture is
metabolized. Consequently, there is a residue of nitrogen mixture
escaping from the loop through the overpressure valve.
-
For
a closed rebreather, oxygen is added to the loop manually, via a CMF
system, or by an electronic manner.
-
In
the case of manual control, oxygen partial pressure is read from a
display, and if necessary added with a manually controlled oxygen valve
(bypass).
-
A
CMF system, sometimes also called a KISS system, determines the dosage
of oxygen by restriction in a piece of stone or hard metal**.
-
Another much used technique is to make use of a needle (metering) valve.
** The orifice is a "synthetic ruby" (actually is
"synthetic corundum", Al2O3, the red variety is
"ruby", the blue variety is "sapphire"), not a natural gem (it would cost
too much to cut a natural gemstone).(Addendum by Gilberto Bonaga, with
thanks!)
The use of corundum is due to the fact that it is the harder material after
the diamond in the mohs scale and it is possible to produce it industrially
at a relative low price.
It is a matter of
fine-tuning in such way that the dosage is slightly under the metabolic
usage of the diver and only manual correction of the oxygen partial pressure
is needed.
In case of an
electronically controlled rebreather, oxygen cells measure the oxygen
pressure and correct this via an automatic injection system so that the
oxygen pressure is close to the ideal oxygen pressure (setpoint) all the
time.
In this article I would
like to further go into the measurements and choice of orifices and the
effect on gas injections during malfunctioning. The link to the active SC
rebreathers, as with MCCC rebreathers, can be made because both systems make
use of a constant mass flow that is injected into the loop.
The question is to
determine how a CMF system works. From Jan
Jahns article we learn the following:
-
To achieve a
constant mass flow in intended depths, the flow has to be sonic, i.e.
with the speed of sound in the critical (narrowest) part of the orifice.
-
The
sonic speed is maintained in depths in which the intermediate pressure
is at least twice as high as the ambient pressure.
(In this example we use Pip/Pamb=2, however there
is a more precise calculation)(1)
-
The
amount of gas that is injected into the loop depends on 3 major factors
-
The
intermediate pressure
-
The
composition of the gas
-
The
diameter of the orifice.
Less relevant
peripheral factors within this calculation are:
In order to understand
the choices of a rebreather design, the following arguments are of interest:
-
How
deep do we want to dive with the SCR system?
-
What
maximum oxygen pressure is acceptable?
-
What
is the maximum oxygen consumption per minute?
CNS and UPTD’s are
valid for all rebreathers and are directly related to the used gas. These
limits are well known.
Let us answer following
queries first.
-
In
case we want to dive to a depth of 40 meters, a minimum IP of 10 bar on
our orifices is obligatory.
-
The
nitrogen fraction in the mix is determined by the accepted maximum
oxygen pressure and EAD.
-
The
maximum oxygen usage depends on the diver’s activity.
It
is important to know the limits of the system. Normally the usage is between
0.3 and 2.5 l/min.
|
|
Taking
the Dolphin as a mathematical model.
According to the
technical manual of the Dolphin, the intermediate pressure has to be set at
16.7 bar +/- 0.5 bar with the 50/50 dosage nozzle and 100 bar in the
cylinder. The orifice will give a flow of 7.3 l/min
For
a correct calculation we have to calculate with the absolute pressure,
meaning the intermediate pressure + atmospheric pressure = 17.7 bar!
At the same
intermediate pressure, the 32% nozzle supplies 15.6 l/min. These data
can directly be read from the technical manual. This enables us to
calculate the required diameter of the orifice.
Here with the general
formula for CMF orifices in a simplified fashion:
Dn= d 2×p0
×A
Where:
Dn
= flow in l/min from the nozzle
d
=
orifice diameter in mm
P0
= intermediate pressure in bar
For
A the following values are applicable
A =
8.65 for pure O2
A
=9.233 - 0.00588×O2 [%] for Nitrox
A =9.71
+0.0011×He[%]2 for Trimix (where the result is the most precise
in a range of 10-50% helium content)
For
example: TMX20/50, d =0.31 mm, p0 = 11 bar :
Dn
=A×p0×d2 =13.17 l/min (exacter
calculation results in 12.91 l/min)
Coming
back to our initial example with the 32% nozzle:
We now
know the following
Dn = 15.6 l/min
P0 =17.7
bar
A = 9.233 – (0.00588x32) = 9.04484
Dn=AxP0xd2
15.6
l/min = 9.04484 x 17.7 bar x d2
15.6 l/min = 160.094 x d2
d2 = 15.6 l/min / 160.094
d = 0.31 mm
The orifice that has an
opening of 0.31 mm at an IP of 17.7 bar.a will deliver a flow of 15.6 l/min.
With this configuration the maximum dive depth will be ((17.7:2)-1)10 = 78.5
meter
The fraction of oxygen that is
inhaled can be calculated using the following formula:
(Dn x Fmix)
– VO2
FiO2 = --------------------------
(Dn-VO2)
Let us first calculate what a
diver inhales in the case of a very high metabolic oxygen usage .
In this case VO2 =
2.5 l/min. The gas mixture still is EAN 32
(15.6 l/min x 0.32 ) – 2.5 l/min
FiO2 = --------------------------------------- =
0.1903
(15.6 l/min – 2.5 l/min)
The inhaled oxygen percentage
is 19.03%
With normal metabolic usage
the diver will inhale VO2 = 1.0 l/min
Then:
(15.6 l/min x 0.32) – 1.0 l/min
FiO2 = --------------------------------------- =
0.2734
(15.6 l/min – 1.0 l/min)
The inhaled oxygen
percentage is then 27.34%
Intermezzo:
These values are of course perfect provided that the IP is maintained and
that the orifices do not clog up. What would be the consequences when the IP
decreases due to incorrect settings? In case the pressure is set to a lower
value of, for instance, 12 bar? Suppose the diver will swim up against the
stream and have a VO2 of 2.5 l/min.
The result can be calculated
as follows :
Dn=A.P0.d2
Dn
= 9.04484 x 12 bar x 0.322
Dn = 11.11 l/min
BTW: at 50 meter the mass flow will decrease because the speed of the gas
will drop down into the subsonic region.
|
|
Example with an IP of 11 bar.a. The CMF flow curve drops starting at 45 meter
|
(11.11 l/min x 0.32) – 2.5 l/min
FiO2 = -------------------------------------------- =
0.123
(11.11 l/min – 2.5 l/min)
The inhaled oxygen
percentage is then only 12.3%! The mixture is hypoxic and the diver will
become unconscious.
Another risk is that the
orifice gets clogged up. Normally in such cases the flow will stop and no
fresh gas will be added to the loop. In case of the Dolphin, Dräger added
extra safety in their dosage device.
The Dräger Dolphin’s dosage
device is fitted with TWO orifices per nozzle. The flow is divided over
these two orifices. In case one gets clogged up the other one will maintain
the gas flow in such way that the diver can safely finish his dive!
|
|
Let’s check this with a
calculation;
Suppose the diver checked his
rebreather and measured the flow. When the IP is carefully adjusted to 17.7
bar.a, the 32% nozzle is connected and the flow test done with 100 bar in the
cylinder, there should be a flow of 15.6 l/min.
In case one of the orifice
gets clogged up the flow will be cut in half, resulting in a 7.8 l/min. When
the diver has a VO2= 1.0 l/min, what will be the oxygen
percentage left to breathe?
(7.8 l/min x 0.32) – 1.0 l/min
FiO2 = -------------------------------------------- =
0.22 %
(7.8 l/min – 1 l/min)
In case of normal metabolic
oxygen use of the diver no problems will occur. The diver will be able to
end his dive breathing 22% of oxygen!
When the diver increases his
efforts by swimming faster or against the current, a
VO2 of 1.5 l/min is a more realistic value
(7.8 l/min x 0.32)
– 1.5 l/min
FiO2 = -------------------------------------------- =
0.16 %
(7.8 l/min – 1.5 l/min)
There is still 16% in the
loop. When the VO2 is further increased there is a risk of
hypoxia.
It is clear that by using
double orifices extra safety is added to the system. However, when the diver
has a high VO2 there is still a big risk. For this reason the
measurement of the PO2 is essential!
We can calculate the inner
diameter of one of the 32% orifices in this way:
IP=P0=17.7 bar
A = 9.233 – (0.00588x32) = 9.04484
Dn=7.8 l/min
Dn=A.P0.d2
7.8 l/min = 9.04484 x 17.7 bar x d2
d = 0.22 mm per robin.
Dosage units of the Dräger Dolphin:
|
|
All right, another example
with the 50% dosage unit:
IP=P0=17.7 bar.a
A = 9.233 – (0.00588x50) = 8.939
Dn=7.3 l/min for two robins. For 1 robin = 3.65 l/min
Dn=A.P0.d2
3.65 l/min = 8.939 x 17.7 bar x d2
3.65 l/min = 158,2203 d2
d=
0.15 mm per nozzle
Calculation of the oxygen fraction with VO2=1 l/min and two
working orifices
(7.3 l/min x 0.50) – 1.0 l/min
FiO2 = -------------------------------------------- =
0.42 %
(7.3 l/min – 1.0 l/min)
There is 42% of oxygen in the loop.
Calculation of the oxygen
fraction with VO2=1 l/min and one working orifice:
(3.65 l/min x 0.50) – 1.0 l/min
FiO2 = -------------------------------------------- =
0.31 %
(3.65 l/min – 1.0 l/min)
There is still 31% of oxygen
in the loop!
Note, although we can
calculate the flow through orifices now, the flow always has to be tested
and measured before the dive. Besides the flow, the partial pressure and MOD
should be carefully monitored. When using semi-closed or closed rebreathers,
a specialized training program has to be followed! The calculation, as used
above, has been the basis for the dosage unit of Jan Jahns. He used
restriction orifices used in the gas-heating industry! Check out his
rebreather here:
..\SemiClosed Rebreathers\Germany\draeger_sms_1.htm
Thanks to:
Caroline Kraaijveld
Jan Jahns
|
|
The following
information is added as a result of the conversation I had with Jahn Jahns
and Dave Thompson.
To get more precise calculations the factor Pip/Pamb=2
can be calculated more accurate.
When Poisson constant of
oxygen k
=cp / cv
gives 1.416 and
pcrit / p0 =[2/(k
+1)] k/(k
-1) =B then B =
[2/(2,416)] 3,403846 = 0,5256.
The values of
the inverted ratio
p0 /pcrit
=1/B = 1/0,5256 = 1,9026
Calculations
should be made with the more precise factor 1,9026
Example: if the IP gauge reads 10 bar the absolute pressure is 11 bar.
This 11 bar.a / 1,9026 gives a IP of 5,78 bar meaning a depth of 47,8
meters!
|
|
After the publication of this article a
conversation between Dave Thompson, Jan Jahns and me (Janwillem Bech)
followed. Since this article is very interesting I added it with permission
to complete this article.
|
|
-
Verzonden: maandag 7 februari
2005 15:05
Hope you are
well. I have been reading your article on CMF stuff and I am not so
sure it is a sonic flow. First the kiss system of adding o2, it is
important that the flow drops as we go deeper otherwise the o2 added
would be too great with increasing depth.
For CMF to work
it has to be 2 x ambient for upstream IP to make the gas sonic,
"maybe" in a kiss valve the IP is typically set at 7bar (old
orifice) or
10-11 bar (new orifice)
If the CMF rule
were to be true then the old orifice would cease to work at 25m and
the new one at 40-50m. This of course does not happen (fortunately)
because what we DON'T want is a cmf flow! We need the o2 flow to
reduce in comparison with depth so the diver continuously receives
the correct number of o2 molecules. This slowing of the flow keeps
the amount of o2 that's is required by the diver constant in the
face of increasing pressure.
Now this is as I
understand it and that of course may be wrong :))))
-
----- Original
Message -----
From: "J.W.Bech"
<jw.bech@quicknet.nl>
Sent: Monday,
February 07, 2005 7:56 PM
Dave T sent me
this answer to the article. The KISS IP
Dave
mentioned
in my opinion is
not correct. The old ones 10 bar IP the new 12 bar IP.
What are your
thoughts about his statement?
-
-----Original
Message-----
Sent: 07
February 2005 21:48
To: jw.bech@quicknet.nl
Cc: dave.thomp@virgin.net
Subject: Re: cmf
Hi, JW and Dave,
I´ll try to
answer:
1. Any gas flow
is sonic (i.e. the speed at the critical section equals the local
sound speed) when the pressure ratio p0/p1>=cca 2 (in our case
p0=IP, p1=ambient press.) "cca" means it depends a little on the gas
whether mono- or biatomic etc. So it has to be sonic when p0=7 bar
and p1= e.g. 3 bar (20 m). At about 25 m the flow starts to cease
with depth (becoming subsonic) and fully ceases when p1=7 bar (60
m). That´s simple.
2. In the KISS
RB the steady flow cca 0,8 lpm should be adjusted giving
hypo-metabolic flow of O2 in depths of use. It is adjusted adjusting
the p0 (=IP). The plug in the water inlet of the 1st stage ensures
the constant absolute IP in any depth. Reading the KISS manual find
in the table of adjustment of flow (for a 0.0035 = cca 0.09 mm diam.
orifice): p0=9bar: 0.5 lpm, 10 bar: 0.6 lpm, 11 bar: 0.7 lpm, 12
bar: 0.8 lpm, 13 bar: 0,9 lpm (calculations using the O2 flow
formula give 0.61, 0.68, 0.75, 0.82 and 0.89 lpm). In the manual 15%
difference is declared as allowable. I haven´t found values for 7
bar but the flow would be insufficient then (about 0.4 lpm).
3. Adjustment of
IP on 12 bar ensures the flow being sonic up to cca 50 m.
Then the flow
slowly decreases (my calculations give 0.82 lpm-sonic-in 50 m, 0.80
in 60 m, 0.75 in 70 m, 0.72 in 80 m, 0.62 in 90 m and then rapid
decrease -in 100 m to 0,45 lpm and of course zero in 110 m). That
simply
means: going
deeper, over the sonic limit, you have to use the manual add valve
more often.
4. Even the
manual states: "The metering orifice flow rate will diminish as the
depth (ambient pressure) increases. How much it decreases depends on
the upstream pressure (regulator pressure setting) versus the
downstream pressure (depth). This is not a fault, it is physics."
(end of citate). The decrease of the flow when crossing certain
critical depth is not intentional, it is simply the matter of p0/p1
< cca 2.
5. Regarding the
"correct number of O2 molecules": remember that the flow
(0.8 lpm) is the
normal pressure volume pro minute in any correct depth (not ambient
pressure related volume) so the number of molecules in the dose is
always the same in any depth (statistically, of course:-).
I hope my answer
is satisfying enough, if not, let me know why, guys.
Wishing clear
water around. Jan.
Well, during
writing this sentences I obtained your next e-mails. But I think
everything is explained in this answer. I’ve got my figures from the
new manual (for 0.08 mm orifice) as you can see. Regarding diving
deeper then to "sonic limit" see the upper calculations for 80
meters: the flow makes always 0.72 lpm which doesn’t differ much
from 0.82 lpm in 50 meters.
Compare also
with the plot of CMF in the article on mass flow.. The drop starts
at critical depth but is not very rapid except at the depths where
the ambient press. draws near to the IP.
Greetings.
-
-
-----Oorspronkelijk bericht-----
-
Verzonden: maandag 7 februari
2005 23:39
-
Aan: 'Jan Jahns'; jw.bech@quicknet.nl
-
Onderwerp: RE:
cmf
-
-
Ok I think I
understand most of that but one thing bothers me if I understand
your reasoning correctly and you were to plot the flow on a graph,
you would have a pretty straight line until the flow went subsonic
then it would curve off?
-
-
If that is
correct that is not what seems to happen in practise
-
-
I notice no
additional need to touch the add button when deep over when
shallow. An example, I need to set my IP to give a flow of 0.68lpm
(this matches my metabolic rate when diving) with this set and at
65m I did one hour on the bottom (the deck of a huge ship in nice
clear water) and I didn’t touch the manual add button once. The IP
set to give this flow is 10bar (according to my gauge) if your
theory/ maths is correct it should have gone sub-sonic at 40m, is
this correct?
-
-
Thank you for
your help and explanations
-
-
-
-----Original Message-----
-
Sent: 07 February 2005 23:35
-
To: 'David Thompson'
-
Cc: jahns@quick.cz
-
Subject: RE: cmf
-
-
Dave; Jan;
-
-
Greetings. Very interesting! Dave, when I look at the curve it
doesn't curve off. It seems almost flat until 80 meter.
-
I think what Jan supposed would be similar to your findings Your
VO2=0,68 and 70 m= 0.72 in 80 m= 0.62.(how many injections would be
needed for the 0,68 & 0,62 difference?) In my opinion that would
explain your findings.
-
-
-
This message reached me too:
-
-
"the orifice is a "sinthetic ruby" (actually is "sinthetic
corundum", Al 2O3, the red variety is "ruby", the blue variety is
"sapphire"), not a natural gem (it would cost too much to cut a
natural gemstone).
-
-
The use of corundum is due to the fact that it is the harder
material after the diamond in the Mohs scale and it is possible to
produce it industrially at a relative low price.
-
"
-
The person telling this is Gilberto Bonaga a geologist.
-
-
OK for now, good night;
-
-
-
-
I do wonder however if its not sonic but just simple physics "double
the pressure, halve the volume" type of stuff, we are just lucky
that it fits perfectly with our metabolic needs!!!
-
-
A couple of engineers I spoke to (who know far more about this than
I ever
-
will) also doubt the sonic flow of this type of gas addition
-
-
They argue that were it sonic you would see a large change in the
gas addition once the 2 x ambient upstream pressure diminished!!!
-
-
Anyway very interesting conversation guys
-
-
Best
-
-
-
-
-----Original Message-----
-
Sent: 08 February 2005 21:27
-
To: David Thompson; jw.bech@quicknet.nl
-
Subject: Re: cmf
-
-
Well, then
-
-
I´ve made some recalculations of the flow depending on depth
according to your low O2 consumption rate. I suppose you used
normal pressure gauge and your reading was 10 atm gauge, i.e. 11 ata
(absolute)
(1).
My calculations give about 0.61 lpm (for 0.08 mm orifice) then, but
let´s take your actually measured value 0.68 lpm as the base for
following calculations.
-
More exactly the calculated critical ratio p0/p1 = IP/(ambient
pressure) is not 2 but 1.9026(2)
and 11/1.9026 gives 5.78 bar of the ambient pressure, i.e.
-
47.8 meters as the maximum depth in which the flow changes from
sonic to subsonic. At this depth the flow rate is always 0.68 lpm
and only then started to diminish. The calculated flow rates are
then: in 50 m the 3 th decadic place does not change giving always
0.68 lpm. At 55 m I obtained
-
0.676 lpm. In 60 m: 0.663 lpm and in 65 m: 0.645 lpm. Seems the
little difference (0.035 lpm) is not serious for your needs, Dave
:-).
-
To calculate the subsonic flow is a little bit more complicated then
the sonic one but is not very difficult and your engineering friends
sure could do it as well.
(3)
I use spreadsheet calculator to do the math but during the long
period of use I´ve built in much balast so it´s readable for me only
now.
-
Regarding your doubts about the sonic speed and precise mass flow in
the sonic region:
-
The gas expands adiabatically into place of lower pressure, cools
down and the speed of sound at this temperature is the critical one
which cannot be overcome -in the critical section of the orifice
(usage of properly formed – so called “Laval – nozzle” can enhance
the speed downstream of the nozzle up to Mach 2). This speed is
maintained then even if the downstream pressure grows - (e.g. in
confined space) but only up to this given by the mentioned ratio
(calculations in the article). Only then it becomes subsonic.
-
The mass flowing out equals then (see Continuity Equ.:)
(cross-section) x
-
(speed) x (critical density). As all 3 variables are constant in the
sonic region also the mass flow is maintained constant.
-
-
-
Excellent, thanks for your help, that makes it more understandable
to my
-
non- engineer mind :))
-
-
Best
-
-
-
-
-
-
13 Februari 2005
-
Oosthuizen Netherlands
-
11:17
-
-
- Jan:
-
- Reading all
this I still have three questions:
-
(Jan Jahns answers follow
directly in colour Red)
-
- So for the
calculations we use the ABSOLUTE pressure and not the measured
pressure on the IP gauge! I.e when a IP is measured of 16,7 bar as
Dräger stated in the technical manual the calculations should be
made with a absolute IP of 17,7 bar? (I
changed the values in the article).
http://www.therebreathersite.nl/01_Informative/orifices_in_rebreathers.htm Or
do they give the gasflow for a measured IP?
-
- I think Dräger IP
values are in bar-gauge for custom convenience, so 1 bar has to be
added to obtain the absolute pressure, really. There is only 1 place
in Draeger manual concerning the IP which I have found to be 240 psi/16.5
bar (not psig/ata!). During the check of your article I hold not in
mind these parameters and I supposed you used absolute pressure. In
my article I always mentioned it as well as in graphs.
- But really again:
Dräger allowed flow tolerances are 11% and the flow using either
values (16.7 bar vs. 17.7 bar) lies in 6 % tolerance (the flow is
proportional to IP). The calculated orifice diam. using 17.7 bar
changes from 0.156 to 0.162 mm. It makes only a little difference,
nevertheless it is methodically wrong if 16,7 is used in the
calculations.
-
-
- You stated that
in this case the
p0/p1 =
IP/(ambient pressure) is not 2 but 1.9026.
The values of the
inverted ratio p0 /pcrit =1/B
lie for the breathing gases in the range 1,90 -2,05, i.e.
p0 /pcrit in average aproximately
equals 2.
- What factor(s)
determin this exact calculation? In other words, how did you
calculate the 1.9026 factor in Dave’s case?
- It´s simple: I took κ = 1.416 for O2
(which I use in my calculation spreadsheet) and inserting it to the
formula you can find 6 rows down I got B = 0,5256 which inverted
gives 1/B = 1.9025875, rounded 1.9026.
-
- Should
the flow velocity be sonic (equal the velocity of sound) we have to
put w2 =c2 and from equations derived
we obtain that for the ratio p/p0 (i.e. for the
ratio of the outlet to inlet pressures) following condition has to
be fulfilled:
-
pcrit
/ p0 =[2/(k
+1]
k/(k
-1)
=B
.
- Here
pcrit is the critical pressure in the critical
section of the nozzle and from the formula follows that if the
outlet pressure p is lower or equal to pcrit
, i.e. if p/p0
£
B
holds in
the critical section the local sonic velocity wcrit
is achieved which can't
be
exceeded increasing the inlet pressure p0.
-
-
- Could you
sent me an example of the subsonic flow calculation, using the
calculated examples you sent to Dave and me.
-
-
The calculated
flow rates are then: in 50 m the 3 th decadic place does not change
giving always 0.68 lpm. At 55 m I obtained
-
0.676 lpm. In 60
m: 0.663 lpm and in 65 m: 0.645 lpm. Seems the little difference
(0.035 lpm)
-
-
As I wrote it´s a problem: I wrote the programs
for myself so there are many odds which currently only I understand.
I have to rewrite some but it takes time. Nevertheless: (I prefer
using bigger Arial to make types in formulae beter readable). Using
asterix for multiplication sign.
-
-
D =
dVn / dt
(normal liters pro minute) = 1/ρn * dM / dt =
1/ρn * S * w * ρ,
-
M
-
mass (kg), ρ – density (kg/m3) in orifice, ρn
– normal density = 1.409 kg/m3 for O2 , S
= π * (d/4)2 –cross section of the orifice
and w – the velocity of gas in the orif. – now subsonic ! It
could be calculated by the help of
formula you can find some rows lower.
Insert IP (absolute) as p0 , ρ0
= ρn * (p0/pn ), ρ = ρ0
* (p/p0)1/κ . Put all variables
in the above bold formula , do some algebra and you have the
right formula into which the correct figures can be inserted.
-
Notice: There are mostly ratios of pressures there, so mostly bars
could be used, only once in the √ you have to use 101300 (Pascals
) for pn . Inserting d in mm you also have
to divide the result by 10002 (=106), then
multiplicate by 1000 (mm = 1 m) and by 60 (seconds in 1 minute).
-
- Jan thanks
for the efforts!
-
- Janwillem Bech
-
- You are welcome, JW
-
Jan Jahns
-
- Inserting
e.g. the value of
k
for biatomic gases gives the value B=0,528.
- The
values of the inverted ratio p0 /pcrit
=1/B lie for the breathing gases in the range 1,90 -2,05,
i.e. p0 /pcrit in average
aproximately equals 2. It means again that the inlet pressure p0
being at least the double of p -the outlet one -the
velocity in the nozzle is sonic. But being lower, e.g. p0
=1.5 p the velocity would be subsonic.
- The
above reasoning implies following: holding the inlet pressure p0
constant (independent of depth) and equal
10 bars
(absolute)
the velocity (i.e.also the gas flow as we´ll see later) is sonic up
to depth in which the ambient pressure is half of it, i.e. up to 40
meters, in which the ambient pressure is 5 bars. The velocity in the
nozzle of air whose inlet temperature has been 0°C is in each
depth up to 5 bars 303 m/s. But going deeper the velocity decreases
onto 230 m/s in 60 m
-
(according
w =Ö{
[2.k/(k
-1)].p0/r0
.[1 -(p/p0) (k
-1)/k]}
), onto
105 m/s in 80 m and the flow would stop in 90 m (where ph
=10 bar ºp0).
-
-
-
|
Jan Jahns and Dave Thompson, thanks for
the value added to this article!
|
|
|